15.1-Inch TFT Color Multi-Parameter Patient Monitor with Expandable IBP & EtCO₂ Options
The 15.1-Inch TFT Color Patient Monitor is an advanced clinical monitoring system designed for adult, pediatric, and neonatal patients. It provides comprehensive monitoring of essential vital signs with advanced ECG analysis, flexible waveform display, and expandable options including IBP, EtCO₂, and thermal printing for enhanced clinical applications.
All Meditech professional healthcare customers can submit equipment service tickets through our Customer Service Portal.
To receive an additional 2 months of extended warranty, please register your products here.For all support resources and contact information, visit our Support hub.
Coming soon.
Discover the advanced capabilities that set our products apart
The 15.1-Inch TFT Color Patient Monitor delivers reliable multi-parameter monitoring with advanced analysis capabilities and flexible expansion options for hospitals, ICU, emergency, and critical care environments
Detailed technical specifications and performance parameters
| Parameter | Specification |
|---|---|
| ECG | |
| Lead Mode | 5 leads (R, L, F, N, C / RA, LA, LL, RL, V) |
| Lead Selection | I, II, III, aVR, aVL, aVF, V |
| Waveform | 2 channels |
| Gain | ×2.5, ×5, ×10, ×20 mm/mV, Auto |
| HR Range | Adult: 15–300 bpm / Neo-Ped: 15–350 bpm |
| HR Accuracy | ±1% or ±1 bpm |
| Resolution | 1 bpm |
| Input Impedance | >5 MΩ |
| CMRR | Monitoring >105 dB / Operation >105 dB / Diagnostic >85 dB |
| Leakage Current | <10 μA |
| ECG Signal Range | ±8 mV (Vp-p) |
| Bandwidth | Surgery: 1–15 Hz / Monitoring: 0.5–35 Hz / Diagnostic: 0.05–100 Hz |
| ST Monitoring | -2.0 ~ +2.0 mV |
| ARR Detection | ASYSTOLE, VFIB/VTAC, PVC, VT, TACHY, BRADY, R ON T, BIGEMINY, TRIGEMINY, etc. |
| Defibrillation Recovery | <3 s |
| RESP | |
| Method | Impedance between RF (RA-LL) |
| Measurement Range | Adult: 0–120 rpm / Neo-Ped: 0–150 rpm |
| Resolution | 1 rpm |
| Accuracy | ±2 rpm |
| Apnea Alarm | 10–40 s |
| NIBP | |
| Method | Oscillometric |
| Modes | Manual / Auto / STAT |
| Pulse Rate Range | 40–240 bpm |
| Alarm Type | SYS / DIA / MEAN |
| Adult Range | SYS 40–270 mmHg, DIA 10–215 mmHg, MEAN 20–235 mmHg |
| Pediatric Range | SYS 40–200 mmHg, DIA 10–150 mmHg, MEAN 20–165 mmHg |
| Neonatal Range | SYS 40–135 mmHg, DIA 10–100 mmHg, MEAN 20–110 mmHg |
| Resolution | 1 mmHg |
| Accuracy | ±5 mmHg mean deviation, ±8 mmHg standard deviation |
| Overpressure Protection | Adult 297±3 mmHg / Pediatric 240±3 mmHg / Neonatal 147±3 mmHg |
| SpO₂ | |
| Measuring Range | 0–100% |
| Resolution | 1% |
| Accuracy | ±2% (70–100%) |
| Update Interval | Approx. 1 sec |
| Pulse Rate Range | 0–254 bpm |
| Pulse Rate Accuracy | ±2 bpm |
| Temperature | |
| Channel | 1 Channel |
| Measuring Range | 0–50°C |
| Resolution | 0.1°C |
| Accuracy | ±0.1°C |
| Update Interval | Approx. 1 sec |
| Response Time | <10 sec |
| IBP (Optional) | |
| Channels | Support IBP measurement |
| Parameters | ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
| ART Range | 0–300 mmHg |
| PA Range | -6–120 mmHg |
| CVP/RAP/LAP/ICP Range | -10–40 mmHg |
| P1/P2 Range | -10–300 mmHg |
| Sensor Sensitivity | 5 μV/V/mmHg |
| Impedance | 300–3000 Ω |
| Resolution | 1 mmHg |
| Accuracy | ±2% or ±1 mmHg |
Compatible accessories to extend functionality