High resolution 12” color TFT display
12 Channel waves.9 Parameters : ECG, Resp, NIBP, SpO2, TEMP,
Light weight, compact and portable
Built-in rechargeable lithium battery
Unique human voice alarm
Powerful storage: Maximum 720 Hour graphic and tabular trends of all
parameters, up to 10000 NIBP review ,200alarm review ,2Hours of all waveforms
System software can be upgraded through network
Arrhythmia analysis and S-T segment analysis
Suitable for adult, paediatric and neonatal patient
Options
Invasive Blood Pressure (IBP)
Carbon dioxide (CO2) Module
Thermal Printer
Touch Screen
ECG
Lead Mode 5 Leads ( R L F N C or RA LA LL RL V)
Lead selection I II III avR avL avF V
Waveform 2 ch
Lead mode 3 Leads ( R L F or RA LA LL)
Lead selection I II III
Waveform 1 ch
Gain ′2.5mm/mV ′5.0mm/mV ′10mm/mV ′20mm/mV
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ±1% or ±1bpm which great
Resolution 1 bpm
Sensitivity > 200 uV P-P
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential ±300mV
Leakage Current < 10 uA
baseline Recovery < 3 s After Defi.
ECG Signal Range ±8 mV (Vp-p)
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
Calibration Signal 1 mV (Vp-p) ±5% Accuracy
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type ASYSTOLE VFIB/VTAC COUPLET BIGEMINY TRIGEMINY R ON T VT>2 PVC TACHY BRADY MISSED BEATS PNP PNC
Alarm Available
Review Available
RESPARATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
base line Impedance Range: 0 – 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy ±2 rpm
Apean Alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual Auto STAT
Measuring Interval in AUTO Mode 1 2 3 4 5 10 15 30 60 90 120 180 240 480 Min
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS DIA MEAN
Measuring and alarm Range Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg
Accuracy Pressure
Maximum Mean error ±5mmHg
Maximum Standard deviation ±8mmHg
Overpressure Protection
Adult Mode 297±3 mmHg
Pediatric Mode 240±3 mmHg
Neonatal Mode 147±3 mmHg
SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy 70% ~ 100% ±2 %
0% ~ 69% unspecified
Actualization interval about 1Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range
0~254bpm
Resolution 1bpm
Accuracy ±2bpm
Alarm Delay 10 Sec.
TEMPERATURE
Channel 2
Measuring and Alarm Range 0 ~ 50 °C
Resolution 0.1°C
Accuracy ±0.1°C
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.
Options: EtCO2 IBP Recorder
Standard:
ECG Cable (5 Leads) ...... 1Pcs
SpO2 Cable (Adult) ........ 1Pcs
TEMP Probe (Adult) ....... 1Pcs
NIBP Cuff (Adult) ............ 1Pcs
NIBP Tube(150cm) ......... 1PcsGround Cable ................ 1Pcs
Power Cable ................... 1Pcs
Cleaning Cloth ................ 1Pcs
Options:
ECG Cable (3 Leads)
ECG Alligator Clips (Metal)
SpO2 Cable (Adult hard shell sensor)
SpO2 Cable (Adult Soft shell sensor)
SpO2 Cable (Neonatal soft sensor)
NIBP Cuff (infant 8 - 14 cm 5/Bag)ECG Alligator Clips (Metal)
SpO2 Cable (Adult hard shell sensor)
SpO2 Cable (Adult Soft shell sensor)
SpO2 Cable (Neonatal soft sensor)
Paper Role
Thermal printer
Patient Monitor Wall mounts (3 Models)
Patient Monitor Trolley (5 Models)
Thermal printer
Patient Monitor Wall mounts (3 Models)
Patient Monitor Trolley (5 Models)
Product Features
High resolution 12” color TFT display
12 Channel waves.9 Parameters : ECG, Resp, NIBP, SpO2, TEMP,
Light weight, compact and portable
Built-in rechargeable lithium battery
Unique human voice alarm
Powerful storage: Maximum 720 Hour graphic and tabular trends of all
parameters, up to 10000 NIBP review ,200alarm review ,2Hours of all waveforms
System software can be upgraded through network
Arrhythmia analysis and S-T segment analysis
Suitable for adult, paediatric and neonatal patient
Options
Invasive Blood Pressure (IBP)
Carbon dioxide (CO2) Module
Thermal Printer
Touch Screen
Product Specifications
Related Accessories
Related items
D-Sense®
Meditech's D-Sense automated external defibrillator simulator AED can simulate a variety of normal and abnormal ECG signals for defibrillator testing.