Defi 5T "AED Trainer" IR remote control
Defi 5T "AED Trainer" IR remote control Note: The remote control battery access is located on the back side of the remote control
Read more
DefiSense Defibrillator Analyzer – Advanced Defibrillator Testing & ECG Simulation System
All Meditech professional healthcare customers can submit equipment service tickets through our Customer Service Portal.
To receive an additional 2 months of extended warranty, please register your products here.For all support resources and contact information, visit our Support hub.
Coming soon.
Discover the advanced capabilities that set our products apart
Detailed technical specifications and performance parameters
| Simulation Mode | Description |
|---|---|
| Sync Time | Synchronization function: ON / OFF |
| Asystole | Simulates cardiac arrest rhythm with no electrical activity |
| VFib 2 Fine | Fine ventricular fibrillation waveform simulation |
| VFib 2 Coarse | Coarse ventricular fibrillation waveform simulation |
| VFib 1 Fine | Fine ventricular fibrillation waveform simulation |
| VFib 1 Coarse | Coarse ventricular fibrillation waveform simulation |
| Monomorphic VTach | Monomorphic ventricular tachycardia simulation |
| Polymorphic VTach 5 | Polymorphic ventricular tachycardia waveform simulation (Level 5) |
| Polymorphic VTach 4 | Polymorphic ventricular tachycardia waveform simulation (Level 4) |
| Polymorphic VTach 3 | Polymorphic ventricular tachycardia waveform simulation (Level 3) |
| Polymorphic VTach 2 | Polymorphic ventricular tachycardia waveform simulation (Level 2) |
| Polymorphic VTach 1 | Polymorphic ventricular tachycardia waveform simulation (Level 1) |
| Normal Sinus Rhythm (NSR) | Standard physiological ECG rhythm simulation |
| Energy Wave | Displays defibrillation energy waveform |
Selectable ECG Signal Amplitudes:
| Amplitude Setting |
|---|
| 5.00 mV |
| 4.50 mV |
| 4.00 mV |
| 3.50 mV |
| 3.00 mV |
| 2.50 mV |
| 2.00 mV |
| 1.50 mV |
| 0.45 mV |
| 0.40 mV |
| 0.35 mV |
| 0.30 mV |
| 0.25 mV |
| 0.20 mV |
| 0.15 mV |
| 0.05 mV |
| Parameter | Specification |
|---|---|
| ECG Display Speed | 25 mm/s |
| Parameter | Specification |
|---|---|
| Maximum Input Voltage | 5000 V peak |
| Internal Load | 50 Ω ±1% |
| Power Supply | 5 V DC, 400 mA (USB Type-B) |
| Defibrillation Energy Measurement Range | 0 – 600 J |
| Voltage Measurement Range | 0 – 5000 V |
| Current Measurement Range | 0 – 100 A |
| Parameter | Specification |
|---|---|
| Energy Accuracy | ±1% of reading ±1 J (0–199.9 J) |
| High Energy Measurement Accuracy | ±4% + 10 µJ |
| Pulse Width Measurement Range | 5 µs – 120 ms |
| Pulse Rate Measurement Range | 5 – 800 ppm |
| Standard Internal Load | 50 Ω |
| Variable Load Range (Optional) | 50 Ω – 1600 Ω (50 Ω steps) |
| Parameter | Specification |
|---|---|
| ECG Output Interface | Standard 5-lead ECG interface |
| ECG Leads | RA/R, LA/L, LL/F, RL/N, V/C |
| Heart Rate Range | 20 – 300 bpm |
| ECG Amplitude Range | 0.05 – 5.00 mV |
| Amplitude Accuracy | ±2% |
| Supported Waveforms | Sine, Square, Triangle, Sawtooth |
| Arrhythmia Simulation | VFib, VTach, Asystole, Normal Sinus Rhythm |
| Parameter | Specification |
|---|---|
| Dimensions | 5.3 × 8.6 × 15 cm |
| Weight | Approximately 0.7 kg |
| Operating Temperature | -15°C to +50°C |
| Storage Temperature | -40°C to +70°C |
| Operating Humidity | 0–95% RH, non-condensing |
| Protection Rating | IP65 |
| Parameter | Specification |
|---|---|
| Safety Classification | Internally powered equipment |
| Applied Part Classification | Type BF, defibrillation-proof |
| Applicable Standard | IEC 60601-2-4 |
Compatible accessories to extend functionality
Coming soon.
Core specifications at a glance
Defi 5T "AED Trainer" IR remote control Note: The remote control battery access is located on the back side of the remote control
Read moreSimulador treino DAE – DEFI 5T TRAINER O DEFI 5T Trainer é uma ferramenta de treino para preparar socorristas para uso desfibrilador AED para tratar uma vítima de paragem cardíaca súbita O simulador fornece uma simulação realística do desfibrilador sem realmente carregar ou descarregar energia eléctrica Instruções de voz, entrega de energia simulada, pausas para RCP, eléctrodos de treino e indicador de status do sistema funcional imitam com precisão a operação do desfibrilador e fornecem uma demonstração realista Inclui controlo remoto para o instrutor dar instruções