    Main stream capnographs VS Side stream capnographs

    June 2026 2 min read

    Types of Capnographs   Main-stream capnographs Side-stream capnographs   CO2 sensor located between endotracheal tube and breathing circuit Sensor is located in the main unit and CO2 is aspirated via a sampling tube connected to a T-piece adapter located between endotracheal tube and breathing circuit

    Types of Capnographs
     

    Main-stream capnographs Side-stream capnographs
    Side-stream CO2

     

    CO2 sensor located between endotracheal tube and breathing circuit Sensor is located in the main unit and CO2 is aspirated via a sampling tube connected to a T-piece adapter located between endotracheal tube and breathing circuit.

     

    Side-stream Capnographs

     

    Advantages Disadvantages
    Easy to connect
    No problems with sterilization
    Can be used in awake patients
    Easy to use when patient is in unusual positions such as in prone position
    Can be used in collaboration with simultaneous oxygen administration via a nasal prong
     
    Delay in recording due to movement of gases from the ET to the unit
    Sampling tube obstruction
    Water vapor pressure changes affect CO2 concentrations
    Pressure drop along the sampling tube affects CO2 measurements
    Deformity of capnograms in children due to dispersion of gases in sampling tubes

     

    Main-stream Capnographs

     

    Advantages Disadvantages
    No sampling tube
    No obstruction
    No affect due to pressure drop
    No affect due to changes in water vapor pressure
    No pollution
    No deformity of capnograms due to non dispersion of gases
    No delay in recording
    Suitable for neonates and children
     
     
     
    Contrary to the earlier versions, the newer sensors are light weight minimizing traction on the endotracheal tube. (see below)
    Long electrical cord, but it is lightweight.
    Facial burns have been reported with earlier versions. This has been eliminated with newer sensors (see below)
    Sensor windows may clog with secretions. However, they can be replaced easily as they are disposable.
    Difficult to use in unusual patient positioning such as in prone positions.
    The newer versions use disposable sensor windows thereby eliminating sterilization problem (see below)
     

     

    Share:

    More from News & Events

    View all articles
    Education
    Education Jun 2026

    Waterproof or water resistant? What’s behind an IP rating?

    Waterproof or water resistant

    Read more 2 min
    Education
    Education Jun 2026

    Reading ECGs (Electrocardiogram)in Dogs

    Reading ECGs (Electrocardiogram) in Dogs  

    Read more 2 min
    Education
    Education Jun 2026

    10 pravil učinkovite prve pomoči

    10 pravil učinkovite prve pomoči

    Read more 2 min