    Index 2 SpO2 Simulator Index 2 Series combines Fingersense our new patented simulated finger with the first comprehensive probe and electronics test to fully analyze pulse oximeters and probes. BIOTEK Index 2 Series SPO2 Simulator,ή ύ,مح

    Index 2 SpO2 Simulator

    Index 2 Series combines Fingersense our new patented simulated finger with the first comprehensive probe and electronics test to fully analyze pulse oximeters and probes. 

    BIOTEK Index 2 Series SPO2 Simulator,προσομοιωτή οξύμετρο,محاكاة مقياس التأكسج,oximeter simulator,血氧饱和度模拟器

    OVERVIEW

    Index 2’s new “Fingersense” assures compatibility with virtually any oximeter on the market today. No hardware or firmware upgrades are necessary. Fingersense incorporates new electronic physiology that instantly recognizes and adapts to all types of oximeter simulation technologies. Use its new user definable “make” feature and pulse oximeter makes that are not included within Index can be verified with acceptable accuracy. With the rapid increase of SPO2 monitors, the new physiology in the electronics help insure that Index 2 will be compatible with your pulse oximeter. 

    Electrical simulation for testing the pulse oximeters alone and probe testing is now available in the “E” version. This allows you to test the oximeter without the probe in-line and all simulations are available. The probe analysis includes:

    LED test: Voltage drop across both red and infrared diodes

    Photodiode test: Detector response to a 1 mA illuminating current (both red and infrared)

    Resistive wire check: Resistances between all wires

    Index 2 lets you simulate arterial oxygen saturation (SPO2) from 35% to 100% oxygenation. Heart rate can also be varied from 30 to 250 BPM. In addition, Plethysmographic waveforms can be set from 0-100% amplitude, simulating the range from no pulse up to a very strong pulse. These basic features allow Index 2 to simulate several different patient conditions such as: 

    Normal Patient

    Patient with a weak pulse

    Geriatric patient

    Obese patient

    Bradycardia

    Tachycardia

    Bio-Tek’s test simulations verify the pulse oximeters SpO2 and heart rate alarms for response time. Use Index 2 for pre-purchase evaluations to determine how well different pulse oximeters respond to artifacts such as motion, ambient sunlight and 50/60 Hz artificial light. 

    In addition, Index 2 comes with full computer control, a built-in RS232, and a printer port. Test results may be transferred to Bio-Tek’s Equipment Management System via OTIS for Windows (OFW) or sent directly to an attached printer for hard-copy results. 

    FEATURES

    Portable

    9 preloaded manufacturers R-Curves

    User definable ‘makes’ for most other manufacturers

    5 downloadable R-curve spaces available >

    Patented physiological finger for a complete SPO2 test

    Electrical probe test and electrical oximeter simulation

    RS232 Port

    APPLICATIONS

    On-going preventative maintenance testing

    Isolate sensor versus electronics problems

    Determine performance variations in less than ideal conditions

    Incoming inspections and repair

    Pre-purchase evaluations

    SPECIFICATIONS

    %O2 

    Range: 35-100%

    Resolution: 1%

    Accuracy: 100-75% ±1%, ±accuracy of the pulse oximeter under test 74-50% ±2% ±accuracy of the pulse oximeter under test Less than 50% unspecified

    Repeatability: ±1 standard deviation

    Rate

    Range: 30-250BPM

    Resolution: 1 BPM

    Accuracy: 1% ±1BPM

    Pulse Amplitude

    Range: 0-100% of nominal pleth amplitude

    Resolution: 1%

    Accuracy: ±1% of setting, nominal pulse

    Display

    2 line x 24 character super twist LCD

    ACCESSORIES

    OTIS for Windows Test Integration Software

    Printer

    Carrying case

    RS232 cable

    Printer cable

    Battery eliminator

    Upgrades

    VERSIONS

    Index 2F

    Optical finger; now with enhanced optics for a more physiological finger

    Index 2E

    Electrical oximeter simulation/testing and electrical probe testing

    Index 2FE

    Both optical finger and electrical simulations/probe testing

    Share:

    More from News & Events

    View all articles
    Education
    Education Jun 2026

    Waterproof or water resistant? What’s behind an IP rating?

    Waterproof or water resistant

    Read more 3 min
    Education
    Education Jun 2026

    Reading ECGs (Electrocardiogram)in Dogs

    Reading ECGs (Electrocardiogram) in Dogs  

    Read more 3 min
    Education
    Education Jun 2026

    10 pravil učinkovite prve pomoči

    10 pravil učinkovite prve pomoči

    Read more 3 min