Education
This guide provides troubleshooting steps for the most common faults of the ECG series devices.
Common faults include:
Device does not turn on
Black screen
Battery cannot charge
Poor display quality
Printing issues
Abnormal or missing ECG waveforms
Note: For all faults, first check whether corresponding cables or connections are loose or disconnected. Reconnect as needed before further troubleshooting.
Press the power button:
Main faults:
A: Faulty key plate or loose/disconnected lines
B: Power supply board fault or loose/disconnected cables
C: If above checks are OK → motherboard may be faulty
Confirm if the issue is the external power supply or the machine (test with another device).
Main faults:
A: Key plate faulty or lines loose/disconnected
B: Power supply board faulty or cables loose/disconnected
After booting, wait ~10 seconds. Press any key:
If the horn sounds → display screen is likely faulty
Symptoms:
Battery cannot charge
Device works on AC but shuts down immediately when unplugged
Battery symbol shows charging but battery level does not increase
Troubleshooting:
Remove the battery and plug in AC. Measure the battery interface with a multimeter:
Normal voltage: 8.2–8.3V → battery is faulty
No or low voltage → power supply board may be faulty
Measure 12V output at power supply board.
No voltage → power supply board is faulty
Voltage OK → motherboard may be faulty
Symptoms: flower screen, missing rows, abnormal colors, white screen
Possible causes: display screen, display driver board, motherboard
Troubleshooting: swap related components; most likely faulty are display or driver board
Causes: printer or motherboard
Check: paper availability, overheating protection, roller installation, printing door pressure
If all normal → likely printer or motherboard issue
Symptoms: incomplete, weak, intermittent, or missing print
Check: printer paper condition and compatibility
If paper is fine → printer or motherboard issue
Symptoms: no waveform, abnormal waveform, or missing lead waveform
Steps:
Clean skin contact area and electrodes with alcohol
Check for lead disconnection; try another lead wire
If unresolved → motherboard may be faulty
Local memory saves up to ~100 cases
For more cases, insert a USB disk:
Steps: Main menu → Working mode → Priority storage → Select USB disk → Press confirm
Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.