- Notebook Ultrasound scanner with 10.1 inch Screen
- B, B/B, B/M, and M image display mode can be chosen.
- Image modification: ¡Á0.8, ¡Á1.0, ¡Á1.2, ¡Á1.3, ¡Á1.5, ¡Á1.6, ¡Á1.8, ¡Á2.0
- Depth advance function is provided.
- support the measurements of distance, circumference, area (polygon approach measuring method and Ellipse measuring method),
- As well as increases the heart rate, slope rate and fetus multi-parameter measurement function.
- 16 kinds of probe position display is provided, 27 kinds of body position are marked.
- Puncture leading, gamma correction, up/down, left/right image reversing, positive/negative
- Real-time clock can show the date and time automatically
- Adjustment of near gain, far gain, total gain can be done alone.
- Multi-stage focus choice and automatic focus function.
- Cine Loop (256 Frames), gamma correction, histogram, etc.
- Four segment real-time dynamic focusing, and the number of focus can be adjusted.
Clinical Applications
Abdominal
Obstetrical/Gynecologist
Fetal
Pediatric
Endovaginal
Endorectal
Technical Specification
Display screen: 10.1¡å
Power Supply: 100 V ~ 240 VAC, 50/60 Hz
Power consumption: ¡Ü 45 VA
Depth of penetration: ¡Ý 180 mm(depend on probes)
Gray scale: 256 levels
Geometric Position Precision error: ¡Ü5£¥(depend on probes)
Average no failure Operation Period: ¡Ý 3000 hours
Battery: 5400 mAh/11.1 V
Available probes
C.D2/3,5 Mhz Convex
MC. D2/3,5 Mhz Micro Convex
L.D27.5 Mhz Linear
T.D2/6.5 Mhz Transvaginal
R.D2/7.5 Mhz Rectal
Probe technical specifications
| Nominal (central)
frequency
|
3.5 MHz
|
| Frequency conversion
range
|
2.5 MHz ~ 5.0 MHz
|
| Type
|
R= 60 mm convex array
|
| Frequency range(MHz)
|
2.5 ¡Ü f < 4.0
|
4.0 ¡Ü f ¡Ü 5.0
|
| Depth of
penetration(mm)
|
¡Ý 180
|
¡Ý 160
|
| Lateral resolution(mm)
|
¡Ü 3
£¨Depth ¡Ü 130£©¡Ü 4
£¨130 < Depth ¡Ü160£©
|
¡Ü
2£¨Depth ¡Ü
60£©
|
| Axial resolution(mm)
|
¡Ü 1(Depth ¡Ü 130)¡Ü 2
(130 < Depth ¡Ü 170)
|
¡Ü 1(Depth ¡Ü 80)
|
| Dead zone(mm)
|
¡Ü 4
|
¡Ü 4
|
| Slice thickness
error, %
|
¡Ü 10
|
¡Ü 10
|
| Geometric Position
Precision error, %
|
¡Ü 5 (Lateral)
¡Ü 5 (Axial)
|
¡Ü 5 (Lateral)
¡Ü 5 (Axial)
|
| Girth, Area
measurement error, %
|
¡Ü 10 (Girth)
¡Ü 15 (Area)
|
¡Ü 10 (Girth)
¡Ü 15 (Area)
|
M mode
| Probe nominal
frequency
|
3.5 MHz
|
| Probe type
|
R= 60 mm convex array
|
| Depth of penetration, mm
|
¡Ý 180
|
| Distance display
error, %
|
¡Ü 10
|
| Time display error, %
|
¡Ü 10
|
Normal Work
Environment
Requirement
for environment
Ambient temperature: 5 ¡æ ~ 40 ¡æ
Relative humidity: ¡Ü 80£¥
Atmospheric pressure: 700 hPa ~ 1060 hPa
|