Measurement Accuracy
SpO2, 70%¡«100% ¡À2%
0%¡«69% undefined
PR ¡À2bpm or ¡À2% (select larger)
Measurement Performance In Weak Filling Condition
pulse-filling ratio : 0.4% SpO2 error ¡À4%,
pulse rate error ¡À2bpm or ¡À2% (select larger)
Optical Sensor
Red light wavelength is 660nm, 6.65mW
Infrared wavelength is 880nm, 6.75mW
Display 2.8'' color LCD Display
NIBP Specifications
Measurement Method Oscillometric method
Working modes Automatic
Measurement Range Pressure adult 0~290mmHg£¨0~38.6kPa£©
pediatric 0~235mmHg£¨0~31.3kPa£©
neonatal 0~140mmHg£¨0~18.6kPa£©
Pulse: 40~240/min
Inflation adult 160mmHg
pediatric 120mmHg
neonatal 70mmHg
Alarm Range adult mode SYS ALM: 40~270 mmHg
DIA ALM: 10~215 mmHg
pediatric mode SYS ALM: 40~200 mmHg
DIA ALM: 10~150 mmHg
neonatal mode SYS ALM: 40~135 mmHg
DIA ALM: 10~100 mmHg
Overpressure protect adult mode 295¡À5mmHg
pediatric mode 240¡À5mmHg
neonatal mode 145¡À5mmHg
Resolution
Pressure 1mmHg
Measurement Accuracy
Cuff Pressure Accuracy ¡À3mmHg
Error The BP Value of the device is equivalence with the measurement value of Stethoscopy. The error meets all the conditions in the YY0667-2008.
Operating Temperature/ Humidity +5oC~40 oC . 15%RH¡«80%RH
Transport and Storage Temperature/Humidity -20oC~+55oC . ¡Ü95%RH
Atmospheric pressure 80kPa~105kPa
Battery 4 "AA" alkaline batteries, AC Adapter separately sold
Main Unit Dimensions 130(L)*110(W)*80mm(H)
Main Unit Weight 300gram
Accessories
Standard :
- Adult Cuff: limb circumference 22-32cm (middle of upper arm )
- SpO2 Probe (Adult)
- Software CD, User Manual, USB data line
Optional :
- Pediatric Cuff: limb circumference 18-26cm (middle of upper arm )
- Neonatal Cuff: limb circumference 10-19cm (middle of upper arm )
- AC Adapter
- Input: AC 100-240V 50-60Hz AC 500mA
- Output: DC 6.0V 1.0A