Defibrillator:
(1) Defibrillator type Manual, Synchronized,
A synchronized,
(2) Output waveform Monophonic
technology
(3) Energy accuracy ∠+1%
(4) Energy selection 0,3,5,7,10,20,30,50,100,200,300,360 joule(nominal at a resistance of
50)
(5) Charging time Maximum 8 seconds
to 360 Joules
(6) Standard Adult/Pediatric
Paddles Reusable external adult
paddles(pediatric paddles integrated)
Patient Monitor:
7''TFT display
ECG:
ECG lead: Ⅰ,Ⅱ,Ⅲ,
a VR, a VL, a VF, V1-V6
Pulse Rate:
Pulse range: 0-300bpm Pulse accuracy: ±2bpm
NIBP: (Defi 9)
Method: automatic oscillation
Working mode: manual/automatic
Measure: adult: 10-250mmHg
Ement: pediatric: 10-200mmHg
Rage: neonatal: 10-135mmHg
TEMP: (Defi 9)
Measurement range: 20-45。C
SpO2: (Defi 9)
Display: SpO2 value, pulse histogram, waveform, pulse
Standard
parameter & configuration
Defi8: ECG .PR .DEFIBRILLATOR
Defi9: ECG .RESP .SPO2 .TEMP .PR .NIBP .DEFIBRILLATOR
Optional parameter
Recorder
Dimensions: L320xW205xH410
Weight: 9.8kg
Technical
specifications
Defibrillator.
(1) Defibrillator
type
Manual, Synchronized, A synchronized,
(2) Output waveform
Monophonic technology /Biphasic
technology (optional)
(3) Energy
accuracy
∠+1%
(4) Energy selection
0,3,5,7,10,20,30,50,100,200,300,360 joule(nominal at a
resistance of 50)
(5) Charging
time
Maximum 8 seconds to 360 Joules
(6) Standard Adult/Pediatric Paddles Reusable external adult
paddles (pediatric paddles
integrated)
Power
requirements.
(a) AC Power
supply
110V, 60 Hz /220V, 50 Hz
(b) Vehicle
Voltage
DC 12V
Internal Battery
Backup
(1) Battery
type
Rechargeable lead-acid battery/ lithium battery (optional)
(2) Charging time
Time-Minimum of 4 hours for full charging
(3)
Battery charging During AC Power Supply operation
of the device.
(4) Charging /Low Battery Indication Red LED
(5) Battery Capacity 100% of the
battery will provide 120 min
of monitoring;
and 30 discharges at 360 joules.
Display:LED
Dimensions 7’diagonal {155mm*88mm}
Type:High resolution true-color graphic LED
Resolution :480X234