| Measuring
range
|
Adult:
30 to 255 mmHg systolic blood pressure
An average of 20 to 235 mmHg pressure
15 to 220 mmHg diastolic blood pressure
Pulse rate of 30 to 240 times / min
Newborn:
30 to 135 mmHg
systolic blood pressure
An average of 20 to 125 mmHg pressure
15 to 110 mmHg diastolic blood pressure
Pulse Rate of 40 to 240 times / min
|
| Measurement
time
|
Adults: up to
more than 120 seconds;
Newborn: up to more than 90 seconds;
|
| Restriction on
software over-voltage
|
Adult:
Less than 290mmHg;
Newborns: no more than 145mmHg;
|
| Protection on
hardware over-voltage
|
Adult: 300 ±10
mmHg;
Newborn: 150 ±4 mmHg;
|
| Measuring
principle
|
Oscillometry
|
| Deflation
approach
|
ladder
Deflated
|
| The initial inflation
pressure value
|
Adult: 150
mmHg;
Newborn: 70 mmHg;
|
| Sensor zero
reset
|
Prior to each
measurement
|
| Resolution
|
1 mmHg
|
| Measurement
accuracy
|
Maximum
average deviation: no more than ± 5 mmHg;
maximum standard deviation: less than 8 mmHg;
|
| Pressure
accuracy instructions
|
± 3 mmHg
|
| Leakage
|
less than 6
mmHg Per minute;
|
| Electrical
performance
|
Voltage: 12
VDC (adapt to the corresponding valve), 5 VDC
Quiescent current: 80 mA
Measurement current: 200 mA
inflated current: 600mA
|
| Environment
|
Operating
Temperature: 0 to 55 Celsius degrees
Storage Temperature: -20 to 65 Celsius degrees
Humidity Range: 0 to 95%, non-condenser
Operation Altitude: -500 m to 4600 meters
Life Span: can operate 100,000 blood pressure measurements
|
| Size
|
12.6 cm x 6.8 cm x 2.8 cm
|
| Weight:
|
180 g
|