Meditech News
Product News
Meditech Library
General info

 

 



 

 




Home > Meditech info > Meditech Library > Auto cpap


Auto cpap


Easy quick one button operation 
Intelligent pressure adjustment 
Automatic pressure tit-ration 
Overheating protection

Meditech iBreath Auto cpap features:

One button easy quick operation

Wide pressure range from 4~20 cm cm H2O can be freely set

The function of Ramp allow patients to fall asleep more comfortably

Wide input range power supply, suitable for international standard

Recording equipment working time and therapy time ,record data up to 3 years

3.5 TFT screen and back light design for easy night operation

Alarm of power -down features to improve security

Automatic leak compensation to ensure that therapy

Automatic altitude pressure compensation

Clock and alarm functions

Integrated designed with heated humidifier

Micro SD card communication function

Overheating protection




Environmental Specifications

 

Working Environment

Storage Environment

Temperature

5 to 35

-20 to 60

Humidity

15% to 95%

(no condensation)

15% to 95% (no condensation)

Atmosphere pressure

77 to 101kPa

Inapplicable

Physical Specifications

Dimension:           255mm*170mm*112mm

Weight:  1.8Kg

Water capacity:    MAX 200ml

Parameter Scope

Pressure :              4~ 20cmH2O (1cmH2O) with a 0.5cmH2O increment

Min pressure:       4~ 20cmH2O (1cmH2O) with a 0.5cmH2O increment

Max pressure:      4~ 20cmH2O (1cmH2O) with a 0.5cmH2O increment

Ramp     0~60min

EPR:       0 level    Close EPR

                1 level    setting pressure -2cmH2O

                2 level    setting pressure -3cmH2O

                3 level    setting pressure -4cmH2O

Humidifier : 0 level    close the humidifier

                  1 level    45

                  2 level    50

                  3 level    55

                  4 level    60

                  5 level    65

System time:        24-hour

Back Light: 30,60,90,120,150,240,330,420,510,600 seconds

Electrical Specifications

 

Humidifier

Doing the test according to YY 0786-2010 standard or equivalent methods.

Output air flow temperature:             <40

Humidity scope:   10-40mg/L

Pressure drop caused by humidifier:              <0.5cmH2O (with the flow rate of 60LPM)

Leaking under maximum working pressure: <25mL/min (Together with the tubing)

Adaptability:          <20mL/kPa (Together with the tubing)

Preheating time:  30 minutes

 

Pressure Accuracy

According to the maximum dynamic pressure changes of ISO 17510-2007 standard.

 

Pressure (cmH2O)

10 BPM

15 BPM

20 BPM

4

0.21

0.5

0.71

8

0.3

0.54

0.75

12

0.39

0.58

0.85

16

0.40

0.65

0.87

20

0.40

0.70

0.97

25

0.53

0.78

1.09

 

MaximumFlow 
According to the maximum flow of ISO 17510-2007 Standard

Set pressure

(cmH2O)

Measured pressure

(cmH2O)

Maximum flow

 (L/min)

4

3

38.2

8

7

38.6

12

12

39.5

16

16

41.1

20

19

41.9

25

24

44.1

 

Power adapter:    Model: DSS-240250 60VA

                                Input: 100-240V, 50-60Hz

                                Output: +24V, 2.5A

Type of protection Against Electric shock:      Class II Equipment

Degree of Protection against Electric shock: Type BF applied part

Degree of Protection against Ingress of Water:            IPX1