Features of Meditech ECG312
- Simultaneously acquisition of 12-lead ECG data, IEC Class I, safety standard Leads: type CF
-
Manual: single channel or three
Auto I: standard three channel
Auto II: standard single channel plus rhythm lead
Auto III: Arrhythmia Analysis
-
Measurement/interpretation and distinguish point printing function under auto mode operation
-
Filters: AC, HUM, DFT
-
Gain: 2.5,5,10,20 and Auto five levels
-
Real-time clock, real-time measurement and presenting of ECG parameter
-
Large LCD (240*320dots) display 12 channel ECG wave forms
-
Battery: high-efficiency lithium battery, at least 2 hours' working time
-
Function of heart rate detection and protection against pulse shock while pacemaker applied
-
Storage of 8 patients' ECG data
-
Power: 220V/110V ( + 10%), 50/60Hz selectable
-
USB interface and RS232 ensures convey of ECG data
-
Option: ECG Software For net-working capacity
Advantages of Meditech ECG312
-
12-lead, 3-channel EKG
-
LCD Display
-
128 Patients' information Storage
-
USB Connection as option
-
RS232 PC Connection as standard Configuration
-
Upgradable for better configuration
-
Recording Paper:63mm*20ml 63mm*30m
-
Standard Accessories: Patient Cable(1 pc), Chest Electrodes(6 pcs), Thermal Paper(1pc), Rechargeable Battery (1pc)
- Gross weight:4.5kg
- Net weight:3.5kg
- Unit size:300*230*61mm
- Packing size:375*295*195mm
|
leads
|
Standard 12 leads
|
|
Acquisition
|
Simultaneous 12 leads
|
|
Resolution
|
18bit/1000Hz
|
|
Operating Mode
|
Manual Mode,3 Auto Modes (with measurement report)
|
|
Filter
|
AC ,EMG Filter
|
|
Drift filter
|
Anti-Drift System
|
|
CMMR
|
>60dB
|
|
>100dB(with AC filter)
|
|
Input Circuit
|
Floating, Protection circuit against Defibrillator effect
|
|
Input Circuit Current
|
<=0.1uA
|
|
Input Impedance
|
>50MQ
|
|
Patient Current Leakage
|
<10uA
|
|
Calibration Voltage
|
1mV+-3%
|
|
Time Constant
|
> 3.2 s
|
|
Frequency Response
|
0.05-150Hz
|
|
Noise Level
|
<=15uVp-p
|
|
Threshold
|
<=20uV
|
|
Paper Speed
|
25, 50mm/s (+-3%)
|
|
Sensitivity
|
Auto,2.5,5, 10, 20 mm/mV
|
|
Recorder
|
High resolution thermal printer
|
|
Paper specification
|
63mmx30m,roll
|
|
LCD display
|
Display menu, operation status and 12 lead ECG waveform
|
|
Safety level
|
IEC60601-1 class I , type CF
|
|
Power Supply
|
AC: 220V/110V+-10%1; 50/60Hz+-1Hz; 30VA
|
|
DC: 14.4V-2000mAH rechargeable lithium battery
|
|
Dimension
|
300mmx230mmx61mm
|
|