Environment conditions
Operation
a) Environment temperature: +5¡æ~+35¡æ
b) Relative humidity: ¡Ü80%
c) Power supply: AC:100~240V,50/60Hz
DC:7.4V, 3700 mAh rechargeable lithium battery
d) Atmospheric pressure: 86kPa~106kPa
Transportation and Storage
a) Environment temperature: -10¡æ~55¡æ
b) Relative humidity: ¡Ü95%
c) Atmospheric pressure: 50kPa~106kPa
Input way: Floating and defibrillation protection
Lead: Standard 12 leads
Patient leak current: <10¦ÌA
Input impedance: ¡Ý5M¦¸
Frequency response: 1Hz~75Hz
Time constant: Time constant>3.2s
CMRR: £¾60dB
EMG interference filter: 25/35Hz(-3dB)
Recording way: Thermal array printing system
Specification of recording paper: 80mm£¨W)¡Á20m(L) high-speed thermal paper
Paper speed:
Auto record: 25mm/s, 50mm/s, error:¡À5%
Rhythm record: 25mm/s, 50mm/s, error:¡À5%
Manual record:5mm/s, 6.25mm/s,10mm/s, 12.5mm/s, 25mm/s, 50mm/s, error:¡À5%.
Sensitivity selections: 2.5,5,10,20,40mm/mV, 5 options ,error:¡À5%. Standard sensitivity is 10mm/mV¡À0.2mm/mV .
Auto record: Record setup according to record format and auto mode, automatically changing leads, measuring and analysing.
Manual record: Record setup according to record format, manually changing leads.
Measurement parameters: HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5) , S(V1) , R(V5)+S(V1)
Product safety type: Class I, Type CF defibrillation proof applied part.
Enduring polarization voltage: ¡À300mV
Noise level: ¡Ü15¦ÌVp-p
Fuse specification: 2 pcs ¦Õ5¡Á20mm AC time lag; T1.6A/250V(Power supply 220V)
Dimension: 315mm(L)¡Á215mm(W)¡Á77mm(H)
Net Weight: 2.25Kg